Technical Specifications
-Defibrillator type Manual, Synchronized, Asynchronized
-Out put waveform Monophasic technology
-Energy accuracy <+1%
-Energy selection 0,3,5,7,10,20,30,50,100,200,300,360 Joule
(Nominal at a resistance of 50)
-Charging time Maximum 8 seconds to 360 Joules
-Paddles Reusable external adult paddles
(Pediatric paddles integrated)
Power Requirements
-AC Power supply 110V, 60 Hz /220V, 50 Hz
-Vehicle Voltage DC 12V
Internal Battery Backup
-Battery type Rechargeable lead-acid battery/lithium battery (optional)
-Charging time Time-Minimum of 4 hours for full charging
-Battery charging During AC Power Supply operation of the device.
-Charging /Low Battery Indication Red LED
-Battery Capacity 100% of the battery will provide 120 mins of monitoring and 30 discharges at 360 joules
Display
-LED Dimensions: 7’diagonal {155mm*88mm}
-Type: High resolution true-color graphic LED
-Resolution: 480X234
ECG
ECG lead: I, II, III, AVR, AVL, AVF, V1-V6, GND
S-T segment analysis: -2.0-2.0mv
Arrhythmia analysis: yes
Protection: With stand 4000vAc/50Hz voltage in isolation and work against electrosurgical interfere and defibrillation
Respiration
Measurement method: Thoracic impedance
NIBP
Working mode: Manual/Automatic/stat (5 minutes)
Method: Automatic oscillation
Cuff type: Standard Adult; optional Pediatric/Neonatal
Temperature
Measurement range 20~45°C
Resolution 0.1°C
Pulse Oximetry (spO2)
Display spo2 value, pulse histogram, waveform, Pulse
Range 0~99%for adult, pediatric and neonatal patients
Probe standard adult finger clip
Optional pediatric Y-type clip and neonatal wrap
Pulse Range
Pulse range: 0-255bpm
Standard Parameter & Configuration
ECG/RESP/SPO2/TEMP/PR/NIBP/DEFIBRILLATOR
Optional
Recorder