Auto Chemistry Analyzer ME-400(CRYSTAL)
Features:
Real-time display reagent remaining volume.
45 Reagent positions.
60 Sample positions (90 sample positions optional).
120 Reaction cuvettes (reusable).
400 tests per hour with STAT function.
Auto 7-step washing system.
One sample probe, one reagent probe, anti-collision
function, liquid level detection function.
Optional built-in sample & reagent barcode reader
Model | ME-400(CRYSTAL) | |
Type | Discrete, fully automatic | |
Throughput | 400T/H | |
Analysis Method | End-point, Fixed-time, Rate, Turbidimetry | |
Certification | CE, FDA, ISO9001, ISO14001, ISO13485 | |
Sample unit | Sample Settings | 60 or 90 sample positions, including calibration, quality control and emergency |
Sample Volume | 2ul ~70ul (0.1ul stepping) | |
Sample Probe | Teflon coating probe, Liquid level sensor, Anti-collision function | |
Probe cleaning | Automatically wash interior and exterior | |
Analysis Items | Analyzing maximally 45 items at the same time | |
Reagent unit | Reagent Position | 45 refrigerated reagent positions |
Reagent Volume | 20~350μl (1μl stepping) | |
Temperature Control | 2℃~8℃cooling for 24 hours | |
Reagent Probe | Teflon coating probe, Liquid level sensor, Anti-collision function | |
Probe Washing | Automatically wash interior and exterior | |
Reaction unit | Temperature Control | 37℃±0.1℃, real-time monitor |
Cuvettes | 120 reusable cuvettes (optical path length 6mm) | |
Mixer Probe | After adding reagent and sample, mix immediately | |
Washing Probe | Automatic water cleaning and drying system | |
System Adjustment | System can record the blank value and get more, accurate results by deductingblank base value of each cuvette | |
Optical System | Light Source | 12V/30W halogen-tungsten lamp, 2000 hours lifetime |
Receiver | 8 precision photo receivers | |
Wavelength Range | 340, 405, 450, 510, 546, 578, 630, 700mm | |
Linear Range | 0.000~3.000Abs | |
Calibration and QC | Calibration | One-point, two-point, multi-point, factor |
Quality Control | Inserting quality control at random, with reaction curve checking each test items clearly | |
Software | Emergency Function | Emergency samples can be placed in the sample tray at any time |
Software | Windows 7/8/10, LIS system available | |
Data Processing | Reaction curve and data display function | |
Others | Power Supply | AC110/220V±10%, 60/50Hz |
Interface | LAN Port | |
Consumption | 500w | |
Package Size(W*D*H) | 1218*958*1431mm | |
Net/Gross Weight | 242kg |
Copyright © Shanghai MedEco Industry Co., Ltd. All Rights Reserved | Sitemap| Powered by |