15" color TFT screen Multi-channel display Big font interface ME9000D
15" color TFT screen Multi-channel display Big font interface
ST and arrhythmia analysis
Drug dose calculation and titration table
OxyCRG dynamic view display
Respiration monitoring with apnea alarms
Alarm review, arrhythmia review
Patient info. Input
Anti-defibrillation design
NIBP over pressure protection
Plug-in rechargeable battery, AC/DC available. Supporting wire or wireless network.
ECG
Lead mode 5 Leads, I, II, III, AVR, AVL, AVF, V
Gain ×2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV,auto
Heart rate Measure range:
Adult 15 ~ 300 bpm
Neonatal/Pediatric 15 ~ 350 bpm
Accuracy ±1%
Resolution 1 bpm
Sensitivity > 200 VμP-P
Differential Input Impedance > 5 M ohm
CMRR Monitor Mode >100 dB
Surgery Mode >100 dB
Diagnostic Mode >90 dB
Electrode offset potential ±300mV
Leakage Current < 10 μA
PACE pulse detect range±2~±700mV
width 0.1~2ms
rise time 10~100 μs
PACE pulse rejection range ±2 ~±700mV
width 0.1~2ms
rise time 10 ~100 μs
Baseline Recovery < 3 s After defibrillation.
Signal Range ±8 mV p-p
Bandwidth Surgery 1 ~ 20 Hz
Monitor 0.5 ~ 40 Hz
Diagnostic 0.05 ~ 130 Hz
Calibration Signal 1 mV p-p, 5% accuracy
ST measurement Range -2.0 ~+2.0 mV
Accuracy-0.8mV~+0.8mV
±0.02mV or±10%, which is greater
Other range: unspecified
Respiration
Method Impedance between RA-LL
Differential Input Impedance >2.5 MΩ
Respiration Impedance Range 0.3~3Ω
Base Impedance Range 200Ω~4000Ω
Bandwidth 0.1 ~ 2.5 Hz
Gain ×0.25,×0.500,×1,×2,×4
Respiration Rate Measurement Range
Adult 0 ~ 120 BrPM
Neonatal / Pediatric 0 ~ 150 BrPM
Resolution 1 BrPM
Accuracy 0~6 BrPM: unspecified
7~150 BrPM:±2 BrPM or±2%
use the greater
Inspiriting current < 300 A RMS max
Apnea Alarm 10 ~ 40 s
NIBP
Method Oscillometry
Measure mode Manual, Auto, Continual
Measure Interval in AUTO Mode 1,2,3,4,5,10,15,30,60,90,120,180,240,480 min
Measure Period in Continual Mode 5 min
Pulse Rate Range 40 ~ 240 bpm
Measure and Alarm Range Adult Mode
SYS 40 ~ 270 mmHg
DIA 10 ~ 215 mmHg
MEAN20 ~ 235 mmHg
Pediatric Mode
SYS 40 ~ 220 mmHg
DIA 10 ~ 150 mmHg
MEAN20 ~ 165 mmHg
Neonatal Mode
SYS 40 ~ 135 mmHg
DIA 10 ~ 100 mmHg
MEAN 20 ~ 110 mmHg
Static pressure accuracy ±3mmHg
Resolution 1mmHg
Accuracy Maximum Mean error ±5mmHg
Maximum Standard deviation 8mmHg
Overpressure Protection Adult 297±3 mmHg
Pediatric 240±3 mmHg
Neonatal 147±3 mmHg
SpO2
Measurement Range 0 ~ 100 %
Resolution 1 %
Accuracy 70% ~ 100% ±2 %
69%
0% ~ 39% unspecified
Pulse Rate Measure and Alarm Range 20~300bpm
Resolution 1bpm
Accuracy ±3bpm
TEMP
Channel 2
Measure and Alarm Range 0 ~ 50℃
Resolution 0.1℃
Accuracy(no sensor) ±0.1(25℃-45℃),0.2±℃(other)
Accuracy(include sensor) ±0.2℃(32℃-42℃)
±0.3(other)
Standard: ECG, Respiration, NIBP, SpO2, Pulse Rate, Temperature
Optional: EtCO2, IBP, Thermal Recorder