Baby incubator ME200(Standard)
Function Features
Microprocessor based servo controlled temperature system
Control modes: Air mode and baby skin mode
Humidity display system
Set temperature, air temperature, baby skin temperature, humidity, timer and heating power are displayed separately by LED
Integral timer to record therapy time
Self - testing function, various failure alarms by audible and visual
Alarm functions:
Power failure,Over temperature,Temperature devitation,Fan failure,Temperature sensor failure
>37℃ temperature set function
Triple protection for over temperature with separate cut off device, more safety system
The inclination of infant bed is adjustable
Single wall hood, 4 operating windows and 2 iris ports,5 infusion seals
Independent locking device for front door
RS232 connector, Oxygen inlet
Package
Each unit is packed in two wooden crate
(control unit and cabinet):112*66*98cm
N.W.:80Kg G.W.:110Kg
Hood:98*63*74cm
N.W.:15Kg G.W.:40Kg
Transport and storage
Environment temperature -10℃~+70℃
Environment relative humidity ≤80%
Atmospheric Pressure 500~1060hpa