Medical Neonatal Incubator Infant Radiant Warmer For Babies
Application of Infant Radiant Warmer:
Suitable for open care, rescue and body temperature adjustment for newborns.
Structure and Composition Of Neonatal Incubator Infant Radiant Warmer:
It consists of a radiation head, a controller, a crib frame, a machine foot, a skin temperature sensor, a tray, an infusion stand, and a phototherapy lamp (optional).
Advantages of Infant Radiant Warmer:
1. Skin temp. and manual contorl by micro-computer
2. Over temperature alarm
3. Available adjustment of radiator and mattress tilting
4. With one illumination light
5. APGAR timer
6. RS-232 connector
7. Low-pressure suction
8. Radiation head that can be adjusted in horizontal angulation.
9. Free-step mattress tilting adjust X-ray tray(option)
Specifications Infant Incubator:
Model | MERW-3000A | MERW-3000B |
Electrical requirement | ~220V50Hz or~230V60Hz | |
Power consumption | ≤700VA | |
Air mode temp. range | 25.0℃-37.0℃ | |
Over temp. alarm | <38.0℃ | |
High mode temp. range | 37.1℃-38.0℃ | |
High mode temp. alarm | <39.0℃ | |
Temp. deviation alarm | ±1.0℃ | |
Skin sensor accuracy | ≤±0.3℃ | |
Mattress temp. uniformity | ≤2.0℃ | |
Heating rate | ≤30min | |
APGAR timing | 1min,5min,10min |
Model | MERW-3000A | MERW-3000B |
Screen | LED | |
Skin temperature senor | ● | ● |
Suction | option | |
O2 bottle | option | |
Electronic lift | option | |
Phototherapy | option | ● |